OAR 333-109-0005
Equipment


(1)

A label bearing essentially the words CAUTION — RADIATION — THIS MACHINE PRODUCES RADIATION WHEN ENERGIZED must be placed near any switch which energizes any portion of the machine. All labels must be the conventional colors (magenta or purple on yellow background) and bear the conventional radiation symbol.

(2)

Any apparatus used in beam alignment procedures must be designed in such a way that excessive radiation will not strike the operator.

(3)

Any switch or device which may cause the radiation machine to produce radiation when actuated must be located on a control panel or console, and must cause a warning light immediately adjacent to such switch or device to light; this light must remain lit when, and only when, the associated control circuit is energized.

(4)

A lock must be provided on the control panel or console that prevents unauthorized operation of the accelerator.
Last Updated

Jun. 8, 2021

Rule 333-109-0005’s source at or​.us