OAR 333-116-0540
Radiation Monitoring Device
(1)
A licensee must have in each teletherapy room a permanent radiation monitor capable of continuously monitoring beam status.(2)
Each radiation monitor must be capable of providing visible evidence of a teletherapy unit malfunction that results in an exposed or partially exposed source. The visible indicator of high radiation levels must be observable by an individual prior to entering the teletherapy room.(3)
Each radiation monitor must be equipped with a backup power supply separate from the power supply to the teletherapy unit. This backup power supply may be a battery system or other type of uninterruptible power supply (UPS).(4)
Each radiation monitor must be checked with a dedicated check source for proper operation each day before the teletherapy unit is used for treatment of patients.(5)
A licensee must maintain a record of the check required by section (4) of this rule until inspection by the Authority. The record must include the date of the check, notation that the monitor indicates when the source is exposed and the initials of the individual who performed the check.(6)
If a radiation monitor is inoperable, the licensee must require any individual entering the teletherapy room to use a survey instrument or audible alarm personal dosimeter to monitor for any malfunction of the source exposure mechanism that may result in an exposed or partially exposed source. The instrument or dosimeter must be checked with a dedicated check source for proper operation at the beginning of each day of use. The licensee must keep a record as described in section (4) of this rule.(7)
If a radiation monitor is inoperable, the licensee must require any individual entering the teletherapy room to use a survey instrument or audible alarm personal dosimeter to monitor for any malfunction of the source exposure mechanism. The instrument or dosimeter must be checked with a dedicated check source for proper operation at the beginning of each day of use. The licensee must keep a record as described in section (5) of this rule.(8)
A licensee must promptly repair or replace the radiation monitor if it is inoperable.
Source:
Rule 333-116-0540 — Radiation Monitoring Device, https://secure.sos.state.or.us/oard/view.action?ruleNumber=333-116-0540
.