OAR 333-123-0060
Facility Design Requirements for Electronic Brachytherapy Devices


(1)

In addition to shielding adequate to meet the requirements of division 120 of this chapter, a treatment room where an electronic brachytherapy device is used shall:

(a)

Prevent simultaneous operation of more than one therapeutic radiation machine in a treatment room if applicable;

(b)

Have a door at any entrance; and

(c)

Permit communication with and visual observation of the patient from the treatment control panel during the irradiation from an electronic brachytherapy device.

(2)

For electronic brachytherapy devices capable of operating at or below 50 kV, radiation shielding for the staff in the treatment room shall be available, either as a portable shield or as localized shielded material around the treatment site.

(3)

For electronic brachytherapy devices capable of operating at greater than 150 kV:

(a)

The control panel shall be located outside the treatment room; and

(b)

Electrical interlocks shall be provided for all doors providing entrance into the treatment room that can:

(A)

Prevent the operator from initiating the treatment cycle if a door remains open;

(B)

Cause the source to be shielded when an entrance door is opened; and

(C)

Prevent the source from being exposed following an interlock interruption until all treatment room entrance doors are closed and the source on-off control is reset at the console.

Source: Rule 333-123-0060 — Facility Design Requirements for Electronic Brachytherapy Devices, https://secure.­sos.­state.­or.­us/oard/view.­action?ruleNumber=333-123-0060.

333–123–0001
Purpose and Scope
333–123–0005
Definitions
333–123–0010
General Requirements, Operating Procedures, Surveys, Dosimetry, and Instrument Calibration for Facilities Using Therapeutic Radiation Machines
333–123–0015
Training and Qualification Requirements for Individuals in the External Beam Radiation Therapy Area
333–123–0020
Reports and Notifications of Unplanned Medical Treatment
333–123–0025
Requirements for Therapeutic Radiation Machines
333–123–0030
Facility Design Requirements for Therapeutic Radiation Machines
333–123–0035
Calibration, Acceptance Testing and Commissioning of Therapeutic Radiation Machines
333–123–0040
Quality Assurance And Safety Checks
333–123–0045
Quality Assurance Program
333–123–0050
Qualified Medical Physicist Support in Facilities Having Therapeutic Radiation Machines with Energies of 1 Mv and Above
333–123–0055
Electronic Brachytherapy
333–123–0060
Facility Design Requirements for Electronic Brachytherapy Devices
333–123–0065
Electrical Safety for Electronic Brachytherapy Devices
333–123–0070
Control Panel Functions
333–123–0075
Timer
333–123–0080
Medical Physicist
333–123–0085
Operating Procedures
333–123–0090
Safety Precautions for Electronic Brachytherapy Devices
333–123–0095
Electronic Brachytherapy Source Calibration Measurements
333–123–0100
Periodic and Day-of-Use Quality Assurance Checks for Electronic Brachytherapy Devices
333–123–0105
Therapy Related Computer Systems
333–123–0110
Training
333–123–0115
Mobile Electronic Brachytherapy Service
333–123–0120
Superficial Electronic Brachytherapy
333–123–0125
Superficial Electronic Brachytherapy Personnel Requirements
333–123–0130
Superficial Electronic Brachytherapy: Refresher Training
333–123–0135
Superficial Electronic Brachytherapy: Activation of the Superficial Electronic Brachytherapy X-ray Tube
333–123–0140
Superficial Electronic Brachytherapy: Facility Design Requirements
333–123–0145
Superficial Electronic Brachytherapy: Calibration, Acceptance Testing and Commissioning
333–123–0150
Superficial Electronic Brachytherapy: Quality Assurance Checks
333–123–0155
Superficial Electronic Brachytherapy: Operating Procedures
Last Updated

Jun. 8, 2021

Rule 333-123-0060’s source at or​.us