OAR 333-123-0105
Therapy Related Computer Systems


(1)

A registrant shall perform acceptance testing on the treatment planning system of electronic brachytherapy-related computer systems in accordance with current published recommendations from a recognized national professional association with expertise in electronic brachytherapy (when available). In the absence of an acceptance testing protocol published by a national professional association, the manufacturer’s acceptance testing protocol shall be followed.

(a)

Acceptance testing shall be performed by, or under the direct supervision of a Qualified Medical Physicist. At a minimum, the acceptance testing shall include, as applicable, verification of:

(A)

The source-specific input parameters required by the dose calculation algorithm;

(B)

The accuracy of dose, dwell time, and treatment time calculations at representative points;

(C)

The accuracy of isodose plots and graphic displays;

(D)

The accuracy of the software used to determine radiation source positions from radiographic images, if applicable; and

(E)

If the treatment-planning system is different from the treatment-delivery system, the accuracy of electronic transfers of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.

(2)

The position indicators in the applicator shall be compared to the actual position of the source or planned swell positions, as appropriate, at the time of commissioning.

(3)

Prior to each patient treatment regimen, the parameters for the treatment shall be evaluated and approved by a Radiation Therapy Physician and a Qualified Medical Physicist for correctness through means independent of that used for determination of the parameters.

Source: Rule 333-123-0105 — Therapy Related Computer Systems, https://secure.­sos.­state.­or.­us/oard/view.­action?ruleNumber=333-123-0105.

333–123–0001
Purpose and Scope
333–123–0005
Definitions
333–123–0010
General Requirements, Operating Procedures, Surveys, Dosimetry, and Instrument Calibration for Facilities Using Therapeutic Radiation Machines
333–123–0015
Training and Qualification Requirements for Individuals in the External Beam Radiation Therapy Area
333–123–0020
Reports and Notifications of Unplanned Medical Treatment
333–123–0025
Requirements for Therapeutic Radiation Machines
333–123–0030
Facility Design Requirements for Therapeutic Radiation Machines
333–123–0035
Calibration, Acceptance Testing and Commissioning of Therapeutic Radiation Machines
333–123–0040
Quality Assurance And Safety Checks
333–123–0045
Quality Assurance Program
333–123–0050
Qualified Medical Physicist Support in Facilities Having Therapeutic Radiation Machines with Energies of 1 Mv and Above
333–123–0055
Electronic Brachytherapy
333–123–0060
Facility Design Requirements for Electronic Brachytherapy Devices
333–123–0065
Electrical Safety for Electronic Brachytherapy Devices
333–123–0070
Control Panel Functions
333–123–0075
Timer
333–123–0080
Medical Physicist
333–123–0085
Operating Procedures
333–123–0090
Safety Precautions for Electronic Brachytherapy Devices
333–123–0095
Electronic Brachytherapy Source Calibration Measurements
333–123–0100
Periodic and Day-of-Use Quality Assurance Checks for Electronic Brachytherapy Devices
333–123–0105
Therapy Related Computer Systems
333–123–0110
Training
333–123–0115
Mobile Electronic Brachytherapy Service
333–123–0120
Superficial Electronic Brachytherapy
333–123–0125
Superficial Electronic Brachytherapy Personnel Requirements
333–123–0130
Superficial Electronic Brachytherapy: Refresher Training
333–123–0135
Superficial Electronic Brachytherapy: Activation of the Superficial Electronic Brachytherapy X-ray Tube
333–123–0140
Superficial Electronic Brachytherapy: Facility Design Requirements
333–123–0145
Superficial Electronic Brachytherapy: Calibration, Acceptance Testing and Commissioning
333–123–0150
Superficial Electronic Brachytherapy: Quality Assurance Checks
333–123–0155
Superficial Electronic Brachytherapy: Operating Procedures
Last Updated

Jun. 8, 2021

Rule 333-123-0105’s source at or​.us