OAR 918-306-0400
Field Evaluation Firm Electrical Equipment Evaluation


The firm shall calibrate, verify and maintain its test and measurement equipment used to conduct any field evaluation. It shall maintain:

(1)

A description of the procedures used in calibrating, verifying and maintaining the test equipment, including as applicable:

(a)

Calibration and verification equipment or services used;

(b)

Reference standards and materials used;

(c)

Measurement assurance, corroborative references, or other programs in which the laboratory participates;

(d)

Specified maintenance practices.

(2)

Calibration and verification records, including as applicable:

(a)

Equipment description or name;

(b)

Name of manufacturer;

(c)

Model, style and serial number, or other identification;

(d)

Equipment variables subject to calibration and verification;

(e)

Statement of the instrument’s allowable error and tolerances of readings;

(f)

Calibration and verification schedules (intervals);

(g)

Dates and results of last calibrations including “as received” results or verifications and schedule of future calibrations or verifications;

(h)

Name of laboratory person or outside contractor providing the calibration or verification service;

(i)

Traceability to National Institute of Standards and Technology or other standard reference authority as required.