OAR 333-122-0110
General Requirements for All RGD: Surveys


(1) All X-ray systems shall have radiation surveys performed, as required by OAR 333-120-0200 (Surveys and Monitoring: General):
(a) Upon installation of the equipment;
(b) Following any change in the initial arrangement, number or type of local components in the system;
(c) Following any maintenance requiring the disassembly or removal of a local component in the system;
(d) During the performance of maintenance and alignment procedures if the procedures require the presence of a primary X-ray beam when any local component in the system is disassembled or removed;
(e) Any time a visual inspection of the local components in the system reveals an abnormal condition; and
(f) Whenever personnel monitoring devices show a significant increase over the previous monitoring period or the readings are approaching the limits specified in OAR 333-120-0100 (Radiation Dose Limits: Occupational Dose Limits For Adults).
(2) Radiation levels must not exceed the limits outlined in OAR 333-120-0180 (Radiation Dose Limits: Dose Limits for Individual Members of the Public).
(3) The registrant must conduct all surveys with a calibrated and operable radiation survey instrument.

Source: Rule 333-122-0110 — General Requirements for All RGD: Surveys, https://secure.­sos.­state.­or.­us/oard/view.­action?ruleNumber=333-122-0110.

333‑122‑0001
Purpose
333‑122‑0003
Scope
333‑122‑0005
Definitions
333‑122‑0010
General Requirements for All RGD: Radiation Safety Officer
333‑122‑0020
General Requirements for All RGD: Safety Device
333‑122‑0030
General Requirements for All RGD: Warning Devices
333‑122‑0040
General Requirements for All RGD: Ports
333‑122‑0050
Exemptions
333‑122‑0055
General Requirements for All RGD: Labeling
333‑122‑0060
General Requirements for All RGD: Shutters
333‑122‑0065
General Requirements for All RGD: X-ray Tube Registration
333‑122‑0070
General Requirements for All RGD: Warning Lights
333‑122‑0076
General Requirements for All RGD: and Radiation Source Housing Requirements
333‑122‑0090
General Requirements for All RGD: Generator Cabinet
333‑122‑0101
General Requirements for All RGD: Radiation Levels
333‑122‑0110
General Requirements for All RGD: Surveys
333‑122‑0120
General Requirements for All RGD: Posting.
333‑122‑0130
General Requirements for all RGD: Operating Requirements
333‑122‑0140
General Requirements for All RGD: Operating and Emergency Procedures
333‑122‑0151
General Requirements for All RGD: Copies of Operating and Emergency Procedures
333‑122‑0160
General Requirements for All RGD: Training
333‑122‑0170
General Requirements for All RGD: Records of Training and Certification
333‑122‑0180
General Requirements for All RGD: Individual Monitoring
333‑122‑0190
General Requirements for All RGD: Dosimetry Individual Monitoring
333‑122‑0201
General Requirements for All RGD: Records of Personnel Monitoring
333‑122‑0210
General Requirements for All RGD: Records of Personnel Monitoring
333‑122‑0226
General Requirements for All RGD: Records of Radiation Survey Instruments
333‑122‑0230
General Requirements for All RGD: Inspection and Maintenance of RGD X-ray Machines, Associated Equipment, and Survey Instruments
333‑122‑0240
Records of Inspection and Maintenance of RGD, Associated Equipment, and Survey
333‑122‑0251
General Requirements for All RGD: Form of Records ​​​​​​​
333‑122‑0260
General Requirements for All RGD: Recordkeeping Requirements for RGD X-Ray
333‑122‑0270
General Requirements for All RGD: Location of Documents and Records
333‑122‑0276
X-ray and Hybrid Gauges: Safety Device
333‑122‑0280
X-ray and Hybrid Gauges: Warning Signs
333‑122‑0290
X-ray and Hybrid Gauges: Security
333‑122‑0301
X-ray and Hybrid Gauges: Surveys
333‑122‑0310
X-ray and Hybrid Gauges: Surveillance
333‑122‑0320
Permanent X-ray Radiography Installations
333‑122‑0326
Permanent X-ray Radiography Installations: Records of Alarm System and Entrance Control Checks
333‑122‑0380
Whole Body Imagers: Appropriate Uses
333‑122‑0600
Notifications
Last Updated

Jun. 8, 2021

Rule 333-122-0110’s source at or​.us