OAR 333-106-0035
General Requirements: Deliberate Exposures Restricted


Persons shall not be exposed to the useful beam except for healing art purposes until the patient has been evaluated, and a medical need for the X-ray/s is determined, and has been authorized by a physician or Dental Professional licensed to practice the healing arts. Any useful diagnostic information obtained from each exposure shall be reviewed by a practitioner of the healing arts. This provision specifically prohibits deliberate exposure for the following purposes:

(1)

Exposure of an individual for training, demonstration or other purposes unless there are also healing arts requirements and proper prescription has been provided.

(2)

Exposure of an individual for the purpose of healing arts screening:

(a)

Any person proposing to conduct a healing arts screening program shall not initiate such a program without prior approval of the Authority;

(b)

When requesting such approval, that person shall submit the following information.

(A)

Name and address of the applicant and, where applicable, the names and addresses of agents within this state;

(B)

Diseases or conditions for which the X-ray examinations are to be used in diagnoses;

(C)

A detailed description of the X-ray examinations proposed in the screening program to include the estimated total radiation dose received by the individual(s) participating in the screening program;

(D)

Description of the population to be examined in the screening program, i.e., age, sex, physical conditions, and other appropriate information;

(E)

An evaluation of any known alternate methods not involving ionizing radiation which could achieve the goals of the screening program and why these methods are not used instead of the X-ray examinations;

(F)

An evaluation by a qualified expert of the X-ray system(s) to be used in the screening program. The evaluation by the qualified expert shall show that such system(s) do satisfy all requirements of these rules;

(G)

A description of the diagnostic film quality control program;

(H)

A copy of the technique chart for the X-ray examination procedures to be used;

(I)

The qualifications of each individual who will be operating the X-ray system(s);

(J)

The qualifications of the individual who will be supervising the operators of the X-ray system(s). The extent of supervision and the method of work performance evaluation shall be specified;

(K)

The name and address of the individual who will interpret the radiograph(s);

(L)

A description of the procedures to be used in advising the individuals screened and their private practitioners of the healing arts of the results of the screening procedure and any further medical needs indicated;

(M)

A description of the procedures for the retention or disposition of the radiographs and other records pertaining to the X-ray examinations.

(3)

If any information submitted to the Authority under subsection (2)(b) changes, the Authority shall be immediately notified.

(4)

Mammography screening shall be exempt from the requirements of section (2) of this rule if the following conditions are met:

(a)

The requirements set forth in OAR 333-106-0700 (Mammography Requirements: Mammography X-Ray Systems Definitions) to 333-106-0750 (Mammography Requirements: Mammography Personnel Qualifications) of these rules are satisfied.

(b)

All other applicable rules are met.

Source: Rule 333-106-0035 — General Requirements: Deliberate Exposures Restricted, https://secure.­sos.­state.­or.­us/oard/view.­action?ruleNumber=333-106-0035.

333‑106‑0001
Purpose and Scope
333‑106‑0005
Definitions
333‑106‑0010
Administrative Controls
333‑106‑0015
General Requirements: Technique Chart
333‑106‑0020
General Requirements: Written Rules and Procedures
333‑106‑0025
General Requirements: Protection of Patients and Personnel
333‑106‑0035
General Requirements: Deliberate Exposures Restricted
333‑106‑0040
General Requirements: Patient Holding and Restraint
333‑106‑0045
General Requirements: Use of Best Procedures and Equipment
333‑106‑0050
General Requirements: Personnel Monitoring
333‑106‑0055
General Requirements: X-ray Operator Training
333‑106‑0060
General Requirements: Radiation Use and Safety Instructor Qualifications
333‑106‑0101
General Requirements: Diagnostic X-ray Systems
333‑106‑0105
General Requirements: Information and Maintenance Record and Associated Information
333‑106‑0110
General Requirements: Plan Review
333‑106‑0120
General Requirements: Information on Radiation Shielding for Plan Reviews — Optional
333‑106‑0130
General Requirements: Design Requirements for an Operator’s Booth
333‑106‑0201
Fluoroscopic X-ray Systems
333‑106‑0205
Fluoroscopic X-ray Systems Requirements: Activation of the Fluoroscopic Tube
333‑106‑0210
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Entrance Exposure Rates
333‑106‑0215
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Barrier Transmitted Radiation Rate Limits
333‑106‑0220
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Indication of Potential and Current
333‑106‑0225
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Source-to-Skin Distance
333‑106‑0230
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Timer
333‑106‑0235
Fluoroscopic X-ray Systems Requirements: Mobile Fluoroscopes
333‑106‑0240
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Control of Scattered Radiation
333‑106‑0245
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Radiation Therapy Simulation Systems
333‑106‑0301
Additional Requirements for Radiographic Machines: Beam Limitation for Radiographic Systems Other Than Fluoroscopic, Dental Intraoral, Veterinary Systems, or Computed Tomography(CT)
333‑106‑0305
Additional Requirements for Radiographic Machines: Radiation Exposure Control Devices
333‑106‑0310
Additional Requirements for Radiographic Machines: Source-to-Skin Distance
333‑106‑0315
Additional Requirements for Radiographic Machines: Exposure Reproducibility
333‑106‑0320
Additional Requirements for Radiographic Machines: Radiation from Capacitor Energy Storage Equipment in Standby Status
333‑106‑0325
Additional Requirements for Radiographic Machines: Intraoral Dental Radiographic Systems
333‑106‑0345
Purpose and Scope
333‑106‑0350
Computed Tomography X-ray Systems: Definitions
333‑106‑0355
CT Equipment Requirements
333‑106‑0360
Computed Tomography X-ray Systems: Facility Design Requirements
333‑106‑0361
Computed Tomography X-ray Systems: Radiation Protection Surveys
333‑106‑0362
Computed Tomography X-ray Systems: Operating Procedures and Conditions of Operation
333‑106‑0363
Computed Tomography X-ray Systems: Quality Control Program
333‑106‑0364
Computed Tomography X-ray Systems: Initial and Annual Qualified CT Medical Physicist Scanner Performance Evaluations
333‑106‑0366
Computed Tomography X-ray Systems: Dose Limits
333‑106‑0367
Computed Tomography X-ray Systems: Records and Report Retention
333‑106‑0368
Computed Tomography X-ray Systems: Qualified CT Medical Physicist
333‑106‑0369
Computed Tomography X-ray Systems: Report and Notification of a CT Medical Event
333‑106‑0370
Computed Tomography X-ray Systems: Operator Requirements
333‑106‑0601
Veterinary X-ray Systems: Veterinary Medicine Radiographic Installations Additional Requirements
333‑106‑0700
Mammography Requirements: Mammography X-Ray Systems Definitions
333‑106‑0710
Mammography Requirements: Equipment Standards
333‑106‑0720
Mammography Requirements: Quality Assurance Program
333‑106‑0730
Mammography Requirements: Additional Requirements
333‑106‑0735
Mammography Requirements: Breast Density Notification
333‑106‑0750
Mammography Requirements: Mammography Personnel Qualifications
Last Updated

Jun. 8, 2021

Rule 333-106-0035’s source at or​.us