OAR 333-106-0210
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Entrance Exposure Rates


(1)

Fluoroscopic equipment manufactured before May 19, 1995 that is provided with Automatic Exposure Rate Control (AERC) shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 roentgens (R) (2.58 mC/kg) per minute, at a point where the center of the useful beam enters the patient, except:

(a)

During the recording of fluoroscopic images; or

(b)

When optional high-level control is provided. When so provided, the equipment shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 5 R (1.29 mC/kg) per minute at a point where the center of the useful beam enters the patient, unless the high-level control is activated. Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

(2)

Fluoroscopic equipment that is not provided with AERC shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 5 R (1.29 mC/kg) per minute at a point where the center of the useful beam enters the patient, except:

(a)

During the recording of fluoroscopic images; or

(b)

When optional high-level control is activated. Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

(3)

Equipment with both an AERC mode and a manual mode. Fluoroscopic equipment that is provided with both an AERC and a manual mode shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 R (2.58 mC/kg) per minute in either mode at a point where the center of the useful beam enters the patient, except:

(a)

During the recording of fluoroscopic images; or

(b)

When the mode or modes have an optional high-level control, in which case that mode or modes shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 5 R (1.29 mC/kg) per minute at a point where the center of the useful beam enters the patient, unless the high-level control is activated. Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

(4)

Exemptions. Fluoroscopic radiation therapy simulation systems are exempt from the requirements set forth in sections (1), (2), and (3) of this rule.

(5)

For fluoroscopic equipment manufactured on and after May 19, 1995, the following requirements will apply:

(a)

Fluoroscopic equipment operable at any combination of tube potential and current that will result in an exposure rate in excess of 5 R (1.29 mC/kg) per minute at a point where the center of the useful beam enters the patient shall be equipped with AERC. Provision for manual selection of the technique factors may be provided.

(b)

Fluoroscopic equipment shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 10 R (2.58 mC/kg) per minute at a point where the center of the useful beam enters the patient except:

(A)

During the recording of fluoroscopic images from an X-ray image-intensifier tube using photographic film or a video camera when the X-ray source is operated in a pulsed mode.

(B)

When an optional high-level control is activated, the equipment shall not be operable at any combination of tube potential and current that will result in an exposure rate in excess of 20 R per minute at a point where the center of the useful beam enters the patient. Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

(6)

Measuring compliance. Compliance with the requirements of this rule shall be determined as follows:

(a)

Movable grids and compression devices shall be removed from the useful beam during the measurement;

(b)

If the source is below the table, exposure rate shall be measured 1 cm above the tabletop or cradle;

(c)

If the source is above the table, the exposure rate shall be measured at 30 cm above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement;

(d)

For a C-arm type of fluoroscope, the exposure rate shall be measured 30 cm from the input surface of the fluoroscopic imaging assembly, with the source positioned at any available SID, provided that the end of the beam-limiting device or spacer is no closer than 30 cm from the input surface of the fluoroscopic imaging assembly;

(e)

For a lateral type fluoroscope, the exposure rate shall be measured at a point 15 cm from the centerline of the X-ray table and in the direction of the X-ray source with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement. If the tabletop is moveable, it shall be positioned as closely as possible to the lateral X-ray source, with the end of the beam-limiting device or spacer no closer than 15 cm to the centerline of the X-ray table.

(7)

Exemptions. Fluoroscopic radiation therapy simulation systems are exempt from the requirement set forth in section (5) of this rule.

(8)

Periodic measurement of entrance exposure rate shall be performed as follows:

(a)

Such measurement shall be made annually or after any maintenance of the system which might affect the exposure rate; and

(b)

Results of these measurements shall be posted where any fluoroscopist may have ready access to such results while using the fluoroscope and in the record required in OAR 333-106-0105 (General Requirements: Information and Maintenance Record and Associated Information)(1)(c). The measurement results shall be stated in roentgens per minute and include the technique factors used in determining such results. The name of the person performing the measurements and the date the measurements were performed shall be included in the results; and

(c)

Personnel monitoring devices may be used to perform the measurements required by subsection (8)(a) of this rule, provided the measurements are made as described in subsection (8)(d) of this rule;

(d)

Conditions of periodic measurement of entrance exposure rate are as follows:

(A)

The measurement shall be made under the conditions that satisfy the requirements of section (6) of this rule; and

(B)

The kVp shall be the kVp typical of clinical use of the X-ray system; and

(C)

The X-ray system(s) that incorporates automatic exposure control shall have sufficient material placed in the useful beam to produce a milliamperage typical of the use of the X-ray system or the worst case; and

(D)

X-ray system(s) that do not incorporate an automatic exposure control shall utilize a milliamperage typical of the clinical use of the X-ray system.
NOTE: Materials should be placed in the useful beam when conducting these periodic measurements to protect the imaging system.

Source: Rule 333-106-0210 — Fluoroscopic X-ray Systems Requirements: Fluoroscopic Entrance Exposure Rates, https://secure.­sos.­state.­or.­us/oard/view.­action?ruleNumber=333-106-0210.

333–106–0001
Purpose and Scope
333–106–0005
Definitions
333–106–0010
Administrative Controls
333–106–0015
General Requirements: Technique Chart
333–106–0020
General Requirements: Written Rules and Procedures
333–106–0025
General Requirements: Protection of Patients and Personnel
333–106–0035
General Requirements: Deliberate Exposures Restricted
333–106–0040
General Requirements: Patient Holding and Restraint
333–106–0045
General Requirements: Use of Best Procedures and Equipment
333–106–0050
General Requirements: Personnel Monitoring
333–106–0055
General Requirements: X-ray Operator Training
333–106–0060
General Requirements: Radiation Use and Safety Instructor Qualifications
333–106–0101
General Requirements: Diagnostic X-ray Systems
333–106–0105
General Requirements: Information and Maintenance Record and Associated Information
333–106–0110
General Requirements: Plan Review
333–106–0120
General Requirements: Information on Radiation Shielding for Plan Reviews — Optional
333–106–0130
General Requirements: Design Requirements for an Operator’s Booth
333–106–0201
Fluoroscopic X-ray Systems
333–106–0205
Fluoroscopic X-ray Systems Requirements: Activation of the Fluoroscopic Tube
333–106–0210
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Entrance Exposure Rates
333–106–0215
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Barrier Transmitted Radiation Rate Limits
333–106–0220
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Indication of Potential and Current
333–106–0225
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Source-to-Skin Distance
333–106–0230
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Timer
333–106–0235
Fluoroscopic X-ray Systems Requirements: Mobile Fluoroscopes
333–106–0240
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Control of Scattered Radiation
333–106–0245
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Radiation Therapy Simulation Systems
333–106–0301
Additional Requirements for Radiographic Machines: Beam Limitation for Radiographic Systems Other Than Fluoroscopic, Dental Intraoral, Veterinary Systems, or Computed Tomography(CT)
333–106–0305
Additional Requirements for Radiographic Machines: Radiation Exposure Control Devices
333–106–0310
Additional Requirements for Radiographic Machines: Source-to-Skin Distance
333–106–0315
Additional Requirements for Radiographic Machines: Exposure Reproducibility
333–106–0320
Additional Requirements for Radiographic Machines: Radiation from Capacitor Energy Storage Equipment in Standby Status
333–106–0325
Additional Requirements for Radiographic Machines: Intraoral Dental Radiographic Systems
333–106–0345
Purpose and Scope
333–106–0350
Computed Tomography X-ray Systems: Definitions
333–106–0355
CT Equipment Requirements
333–106–0360
Computed Tomography X-ray Systems: Facility Design Requirements
333–106–0361
Computed Tomography X-ray Systems: Radiation Protection Surveys
333–106–0362
Computed Tomography X-ray Systems: Operating Procedures and Conditions of Operation
333–106–0363
Computed Tomography X-ray Systems: Quality Control Program
333–106–0364
Computed Tomography X-ray Systems: Initial and Annual Qualified CT Medical Physicist Scanner Performance Evaluations
333–106–0366
Computed Tomography X-ray Systems: Dose Limits
333–106–0367
Computed Tomography X-ray Systems: Records and Report Retention
333–106–0368
Computed Tomography X-ray Systems: Qualified CT Medical Physicist
333–106–0369
Computed Tomography X-ray Systems: Report and Notification of a CT Medical Event
333–106–0370
Computed Tomography X-ray Systems: Operator Requirements
333–106–0601
Veterinary X-ray Systems: Veterinary Medicine Radiographic Installations Additional Requirements
333–106–0700
Mammography Requirements: Mammography X-Ray Systems Definitions
333–106–0710
Mammography Requirements: Equipment Standards
333–106–0720
Mammography Requirements: Quality Assurance Program
333–106–0730
Mammography Requirements: Additional Requirements
333–106–0735
Mammography Requirements: Breast Density Notification
333–106–0750
Mammography Requirements: Mammography Personnel Qualifications
Last Updated

Jun. 8, 2021

Rule 333-106-0210’s source at or​.us