Rule Rule 333-106-0230
Fluoroscopic X-ray Systems Requirements: Fluoroscopic Timer


(1)

Means shall be provided to present the cumulative on-time of the fluoroscopic tube.

(2)

The maximum cumulative time of the timing device shall not exceed five minutes without resetting.

(3)

A signal audible to the fluoroscopist, or the appropriate operator, shall indicate the completion of any preset cumulative on-time; or if no audible signal is provided, the exposure shall terminate.